The SCU286 has a system for direct measurement of the high voltage delivered by the X-Ray generator to guarantee the accuracy of the radiological parameters. Based on this data the control system maintains the stability of mA and kilovolts to within ± 0.5% in any selection range. SITEX & XS units offer a totally constant quality of exposures, as they are virtually exempt from fluctuations in the power supply. OPERATOR INTERFACE The user-friendly high strength polyester front panel is equipped with a sensory-type 16-key keypad, a high brightness display and an anti-reflection screen. The SCU286 also includes a programmable access codes system that features three security levels (supervisor, operator and service engineer) which limit the access to certain functions in relation to the security level selected. DATABASE First and foremost, our SCU286 control unit is intended to be user-friendly. In order to reduce the number of test shots to a minimum

  • Control boxes - electric
  • Controlunit Sitex
  • Sitexs Controlunit
  • Measurement Sitexs

Product characteristics

kV, mA and time setting steps kV, mA, sec
1.0, 0.1, 1.0
Exposure time range min., sec
1 sec to 99 min 59 sec
Direct kV and mA measurement (Control accuracy) %
Yes / +0.5 / -0.5
Simultaneus display of measurements and settings
Yes
Pilot light indicator (Power ON, X-ray ON, Securities)
3
Independant START and STOP buttons
Yes
Two position safety key rotary switch
X-ray ON / STANDBY
Two position main rotary switch
Mains ON / OFF
High brightness graphic Vacuum Fluorescent Displays (VFD)
2 x 20 with anti-reflection screen
Preheating is function of selected kVs (kV max by default)
Automatic
Exposure time calculator (material, FFD, film, density, kV, mA, time)
Optional
User-defined data enabling exposure time calculation
Optional
Type of power supply
Mains, Power generator
Supply voltage range – SELECTOR FREE – Auto resettable fuses VAC
From 198 to 264
Supply frequency range – SELECTOR FREE Hz
From 45 to 66
Minimum Power Factor %
> 0.7
Max RMS mains current (300 kV / 6 mA - SITEX D3006) A
12 at 220 VAC
Storage ambient temperature range °C
-40 to +80
Working ambient temperature range °C
-25 to +55
Protection Class
IP65
Weight Kg
14